Hi,are you ready?
准备好开始了吗?
那就与我们取得联系吧
有一个医疗器械项目想和我们谈谈吗?您可以填写右边的表格,让我们了解您的项目需求,这是一个良好的开始,我们将会尽快与你取得联系。当然也欢迎您给我们写信或是打电话,让我们听到你的声音!

格慧泰福(GHTF) | 高端医疗器械临床注册专家

24小时免费咨询热线:

400-9905-168

填写您的项目信息

填写完表单后,请点击以下任意一种沟通方式:

QQ咨询 QQ咨询 在线商桥

我们的服务
Australian Medical Devices   Guidance Document       

DISCLAIMER

This document is provided for guidance only. The classification examples provided might vary if

a manufacturer assigns a different intended purpose to that used for the examples. Manufacturers

and sponsors must use the Classification Rules, as contained in the Therapeutic Goods (Medical

Devices) Regulations 2002, to classify their medical devices.

Please refer to the Therapeutic Goods Act, 1989 as amended by the Therapeutic Goods

Amendment (Medical Devices) Bill, 2002 and the Therapeutic Goods (Medical Devices)

Regulations, 2002 for legislative requirements.

FURTHER INFORMATION

The Medical Devices Information Unit of the Office of Devices, Blood and Tissues of the

Therapeutic Goods Administration (TGA) can be contacted by:

Telephone:

Facsimile:

Email:

Website:

Mail:

1800 141 144

(02) 6232 8299

cab.medical.device.information@health.gov.au

http://www.tga.gov.au/devices/devices.htm

PO Box 100

Woden

ACT 2606

©

Commonwealth of Australia 2005

2


AMENDMENT SCHEDULE

Version

Number

  Date of

Amendment

Summary of Amendments

3


CONTENTS

Disclaimer ..............................................................................................................................................2

Further Information................................................................................................................................2

Amendment Schedule............................................................................................................................3

Introduction............................................................................................................................................6

Purpose ...................................................................................................................................................6

Background ............................................................................................................................................6

The classes of medical devices..............................................................................................................6

The classification rules ..........................................................................................................................7

Classifying a medical device.................................................................................................................7

Deciding what classification rules apply.......................................................................................7

Important points to remember .......................................................................................................8

Flowchart 1- deciding what classification rules apply .................................................................9

FLOWCHART 2 CLASSIFICATION RULES - NON INVASIVE DEVICES...........................10

RULE 2 NON INVASIVE MEDICAL DEVICES...................................................................11

Rule 2.1 Non-invasive medical devices – general ......................................................................11

Rule 2. 2 Non-invasive devices intended to channel or store blood, etc ...................................11

Rule 2.3 Non-invasive devices intended to modify the biological or chemical composition of

blood, etc.......................................................................................................................................12

Rule 2.4 Non-invasive devices intended to have contact with injured skin ..............................12

FLOWCHART 3 CLASSIFICATION RULES - INVASIVE DEVICES.....................................13

FLOWCHART 3A CLASSIFICATION RULES - INVASIVE DEVICES..................................14

RULE 3 INVASIVE MEDICAL DEVICES.............................................................................15

Rule 3.1 Invasive devices intended to be used to penetrate body orifices.................................15

Rule 3.2 Surgically invasive devices intended for transient use ................................................15

Rule 3.3 Surgically invasive devices intended for short-term use .............................................16

Rule 3.4 Surgically invasive devices for long-term use and implantable devices ....................17

FLOWCHART 4 CLASSIFICATION RULES - ACTIVE DEVICES ..........................................18

RULE 4 ACTIVE MEDICAL DEVICES .................................................................................19

Rule 4.1 Active medical devices – general .................................................................................19

Rule 4.2 Active medical devices for therapy ..............................................................................19

Rule 4.3 Active medical devices for diagnosis ...........................................................................20

Rule 4.4 Active medical devices intended to administer or remove medicines, etc from a

patient’s body................................................................................................................................20

FLOWCHART 5 CLASSIFICATION RULES –Special rules........................................................21

RULE 5 SPECIAL RULES ........................................................................................................22

Rule 5.1 Devices incorporating a medicine.................................................................................22

Rule 5.2 Devices for contraception or prevention of sexually transmitted diseases .................22

4


Rule 5.3 Devices intended for disinfecting, cleaning, rinsing etc ..............................................22

Rule 5.4 Non-active devices intended to record X-ray diagnostic images................................22

Rule 5.5 Devices containing non-viable animal tissues or derivatives......................................22

Rule 5.6 Devices that are blood bags...........................................................................................23

Rule 5.7 Active implantable medical devices .............................................................................23

Rule 5.8 Medical devices for export only ...................................................................................23

Rule 5.9 Devices that are mammary implants ............................................................................23

APPENDIX A - Definitions..............................................................................................................24

Accessory ...................................................................................................................................24

Active implantable medical device ...........................................................................................24

Active medical device ................................................................................................................24

Active medical device for diagnosis ..........................................................................................24

Active medical device for therapy .............................................................................................24

Ancilliary action .........................................................................................................................25

Biological effect .........................................................................................................................25

Body orifice .................................................................................................................................25

Central circulatory system ..........................................................................................................25

Central nervous system................................................................................................................25

Chemical change ........................................................................................................................25

Direct diagnosis ...........................................................................................................................26

Duration ......................................................................................................................................26

Implantable medical device .......................................................................................................26

Integral ........................................................................................................................................26

Intended purpose ........................................................................................................................26

Invasive medical device ..............................................................................................................26

Measuring function ......................................................................................................................26

Medical device ...........................................................................................................................27

Potentially hazardous .................................................................................................................27

Reusable surgical instrument .....................................................................................................27

Surgically invasive medical device ............................................................................................27

Vital physiological process ........................................................................................................28

Wholly or mainly absorbed .........................................................................................................28

5


INTRODUCTION

This guidance document is one of a series produced to assist a wide-ranging audience to

understand the regulatory system for medical devices in Australia that commenced on 4 October

2002. The system has been established by the Therapeutic Goods Act, 1989 and the Therapeutic

Goods (Medical Devices) Regulations, 2002.

Although each guidance document has been developed to provide information about particular

aspect of the medical devices regulatory system in Australia, it is expected that a certain amount of

cross-referencing to other documents in the series will be inevitable.

PURPOSE

The purpose of this document is to provide guidance on the classification rules for medical devices

and relates to the following segments of the medical device legislation.

• Section 41BD of the Therapeutic Goods Act 1989 (the Act)

• Regulation 3.2 of the Therapeutic Goods (Medical Devices) Regulations 2002 (MD

     Regulations).

• Schedule 2 of the Therapeutic Goods (Medical Devices) Regulations 2002.

• Dictionary – Therapeutic Goods (Medical Devices) Regulations 2002.

BACKGROUND

The risks associated with using medical devices can range from little or no risk to patients and

users to significant potential risks inherent in the type of device. The level of premarket

intervention by the regulator is proportional to the level of potential risk and is established through

a classification system based on that potential risk.

The classification risk is determined from:

! the manufacturer’s intended purpose for the medical device, and

! a set of classification rules. These rules will classify medical devices into one of five classes of

   medical devices.

THE CLASSES OF MEDICAL DEVICES

The five classes of medical device based upon the classification rules are:

! Class I- for low risk medical devices

              - includes low risk devices that are sterile and /or have a measuring function

! Class IIa - for low-medium risk medical devices,

! Class IIb - for medium–high risk medical devices,

! Class III - for high risk medical devices, and

! AIMD- for Active Implantable Medical Devices (are treated as Class III medical devices).

6


THE CLASSIFICATION RULES

Schedule 2 of the MD Regulations describes the classification rules that apply to:

! Non-invasive medical devices (Rule 2):

   •General (Rule 2.1),

   •Device to channel or store blood (Rule 2.2),

   •Device to modify biological or chemical composition of blood (Rule 2.3), and

   •Device to contact with injured skin (Rule 2.4).

!

Invasive medical devices (Rule 3):

•Device used to penetrate body orifices (Rule 3.1),

•Surgically invasive device for transient use (Rule 3.2),

•Surgically invasive device for short-term use (Rule 3.3), and

•Surgically invasive device for long-term use and implantable devices (Rule 3.4).

Active medical devices (Rule 4):

•General (Rule 4.1),

•Active device for therapy (Rule 4.2),

•Active device for diagnosis (Rule 4.3), and

•Active devices to administer or remove medicines from a patient’s body (Rule 4.4).

Special rules for medical devices:

•Device incorporating a medicine (Rule 5.1),

•Device for contraception or prevention of sexually transmitted disease (Rule 5.2),

•Device for disinfecting, cleaning another medical device (Rule 5.3),

•Non-active device to record X-ray diagnostic images (Rule 5.4),

•Device containing animal tissues, cells or other substances, that will need to be rendered

     non-viable to mitigate any infectivity risk or microbial or recombinant tissues, cells or

     other substances (Rule 5.5),

•Device that is a blood bag (Rue 5.6),

•Active implantable device (Rule 5.7),

•Device for export only (Rule 5.8), and

•Device that is a mammary implant (Rule 5.9).

!

!

CLASSIFYING A MEDICAL DEVICE

DECIDING WHAT CLASSIFICATION RULES APPLY

The following decisions need to be made before applying the classification rules:

! Is the product a medical device? Check the definition of a medical device on page 27,

! Do the special classification rules (Schedule 2, Part 5 of the MD Regulations) apply? See

   Flowchart 5- Special rules on page 21,

! Is the device intended to be non-invasive? See Flowchart 2 on page 10,

   (NB- If the device is sterile or has a measuring function conformity evidence will be needed),

! Is the device intended to be invasive through a body orifice? See Flowchart 3 on page 13,

! Is the device intended to be surgically invasive? See Flowchart 3A on page 14, or

! Is the device an active medical device? See Flowchart 4 on page 18.

7


IMPORTANT POINTS TO REMEMBER

!

!

The manufacturer is responsible for determining the class of a device.

The classification rules are based on the manufacturer’s intended purpose, taking into account

the mechanism of action, and in some cases, more than one rule can apply. If this happens, the

higher classification applies with the exception of medical devices for export only (Rule 5.8)

which are classified as Class I.

All the relevant classification rules must be considered to determine the class of the medical

device.

The classification must be consistent with the information accompanying the medical device

including the label, directions for use, brochures and operating manuals.

If the intended purpose is not clear in the information, the TGA will assume an intended

purpose consistent with the purpose generally accepted in clinical practice.

If the device is to be used in combination with another medical device, the classification rules

must be applied separately to each device.

Accessories are classified separate to the medical device they are used with.

The duration for use must be specified for all invasive medical devices.

If the device is not for use in a specific part of the body, its classification is based on the most

critical specified use.

Software intended to drive or influence the use of a medical device falls under the same

classification as the medical device.

Based on the intended purpose, software may be a medical device in its own right.

!

!

!

!

!

!

!

!

!

8



更多
收起
    服务流程资料添加中....
    官方收费资料添加中....
    基础法规资料添加中....
    相关服务资料添加中....

    澳大利亚TGA

    Australian Medical Devices   Guidance Document       

    澳大利益TGA认证介绍

    医疗器械澳大利亚市场介绍

全国服务热线:

400-9905-168

关注我们

粤ICP备13059948号 广州格慧泰福(GHTF)生物科技有限公司 版权所有